Precision laboratory balance
Measures specimen gravitational mass M with a resolution of 0.01 g.
Determine the density of pure copper specimens by linear regression of mass against volume using precision balance and displacement measurements. Explore a virtual apparatus, collect readings and analyse a graph.
Ready for a virtual observation.
Adjust the apparatus, then take a reading.
Readings are generated by an ideal model. The diagram is schematic and is not to scale.
The simulation assumes an ideal homogeneous copper specimen without internal casting porosity, oxide films, or surface contamination. Readings in this virtual module reflect ideal copper (ρ = 8960 kg m⁻³) with standard rounding to 0.01 g on balance mass and 0.1 cm³ on volume.
Measures specimen gravitational mass M with a resolution of 0.01 g.
Measures volume of regular or irregular solids by water displacement with anti-meniscus correction.
Measures diameter and height of cylindrical specimens to calculate volume independently.
Provides homogeneous test cylinders of graded volumes from 5.0 cm³ to 60.0 cm³.
Determine the density of pure copper specimens by linear regression of mass against volume using precision balance and displacement measurements. Suggested independent settings: 10.0, 20.0, 30.0, 40.0, 50.0 cm³.
Release the balance beam arrest and adjust the counterpoise thumb screw until the pointer oscillates symmetrically about the central null mark.
Zero error in a mechanical or electronic balance introduces an additive constant to every mass measurement, distorting the y-intercept.
Measure the height and diameter of each cylindrical specimen at three distinct orientations using a vernier caliper to verify geometric regularity.
Averaging multiple caliper readings across different diameters detects taper, ovality, or surface burrs before immersion.
Fill a graduated cylinder with distilled water to a reference level, tilt slightly to avoid air entrapment, lower the dry specimen gently on a fine thread, and record the new meniscus level.
Surface splashing, droplet loss, or air bubbles adhering to the specimen surface cause systematic volume errors.
Remove the specimen from the water bath and blot it dry with lint-free absorbent paper before placing it on the balance pan.
Residual surface water droplets add false mass to the specimen, biasing the measured density upward.
Place the dry specimen at the center of the pan, slide the poise weights along the beams until equilibrium is restored, and read the total mass to the nearest 0.01 g.
Off-center pan loading creates knife-edge friction and torque, reducing measurement repeatability.
Proceed systematically through the graded series of specimen volumes, logging volume V and mass M for linear regression analysis.
Evaluating multiple independent volumes across a wide dynamic range minimizes leverage bias in the slope determination.
Points are your observations. The fit uses SI x and y values before display scaling.
| # | Specimen volume V (cm³) | Specimen mass M (g) | Volume V (10⁻⁶ m³) | Mass M (kg) | Remove |
|---|
No readings yet. Take a virtual measurement and record it.