Chemistry / Polarography & Ilkovič Equation
FREE TO EXPLORE
LABORATORY 38 · NATURAL SCIENCES

Polarography & Ilkovič Equation lab report

A worked polarography & ilkovič equation lab report with gravimetric data tables, linear regression graphs, calculations, uncertainty and conclusions.

WORKED EXAMPLE · QUANTITATIVE OBSERVATIONS

A complete example laboratory report.

Simulated demonstration data.

This report illustrates formal analytical gravimetry, volumetric calibrations, and linear regression analysis. Replace its observations with your own experimental telemetry when preparing a practical submission.

Student: __________________Student ID: __________________Course: __________________Date: __________________Partner: __________________Instructor: __________________

1. Title and aim

Classical Polarography, Diffusion Currents, and Heavy Metal Speciation (Ilkovič Equation). Determine the aqueous diffusion coefficient of the cadmium(II) ion (Cd²⁺) and verify the Ilkovič equation in classical DC polarography using a dropping mercury electrode (DME) across concentrations from 0.200 mM to 2.500 mM in 0.100 M KCl supporting electrolyte.

2. Abstract

The polarographic reduction of cadmium(II) ions in 0.1000 M potassium chloride supporting electrolyte was investigated across the concentration range 0.200 mM to 2.500 mM at 25.00 ± 0.02 °C using classical DC polarography at a calibrated dropping mercury electrode (DME) versus a saturated calomel electrode (SCE). An inert surfactant, Triton X-100 (0.002% w/v), was incorporated to suppress polarographic streaming maxima of the first kind. Linear potential ramps (-0.200 V to -1.000 V vs SCE at 2.0 mV·s⁻¹) yielded smooth, highly reproducible sigmoidal polarographic waves with an invariant half-wave potential E_1/2 = -0.600 ± 0.002 V vs SCE, verifying reversible two-electron reduction into the expanding mercury drop to form cadmium amalgam. Mean limiting diffusion currents i_d increased strictly linearly from 1.27 μA at 0.200 mM to 15.93 μA at 2.500 mM. Ordinary least-squares regression of i_d against cadmium concentration c yielded an empirical slope m = 6.3719 μA·mM⁻¹ with an exceptional coefficient of determination R² = 1.000000 and an intercept b = -6.5672e-04 μA. Using the calibrated capillary characteristics (m_Hg = 2.000 mg·s⁻¹, t_drop = 3.500 s), application of the Ilkovič equation extracted an aqueous cadmium(II) diffusion coefficient D = (7.20 ± 0.06) × 10⁻⁶ cm²·s⁻¹, demonstrating extraordinary concordance with the authoritative electroanalytical literature benchmark (7.20 × 10⁻⁶ cm²·s⁻¹) with a percentage difference of only 0.0102%.

3. Apparatus and fixed settings

4. Theory and graph choice

Classical polarography, invented by Nobel laureate Jaroslav Heyrovský in 1922, is the foundational technique of modern voltammetry. Utilizing a dropping mercury electrode (DME)—in which a continually renewed, atomically pristine spherical mercury droplet grows and detaches every few seconds under constant hydrostatic pressure—polarography eliminates solid-state electrode fouling and memory effects. In an unstirred deoxygenated electrolyte containing 0.100 M KCl and trace Triton X-100 maxima suppressor, the reversible reduction of cadmium(II) ions (Cd²⁺ + 2e⁻ + Hg ⇌ Cd(Hg)) produces a classic sigmoidal polarographic wave. The limiting diffusion current plateau i_d scales strictly linearly with analyte concentration according to the Ilkovič equation. By measuring diffusion currents across cadmium concentrations from 0.200 mM to 2.500 mM at 25.0 °C, the aqueous diffusion coefficient D of Cd²⁺ is determined with exceptional sub-0.5% precision.

\bar{i}_d = 607 \cdot n \cdot D^{1/2} \cdot m^{2/3} \cdot t^{1/6} \cdot c \implies \bar{i}_d = 6.372 \cdot c \implies D = 7.20 \times 10^{-6}\text{ cm}^2\cdot\text{s}^{-1}

The primary analytical visualization presents the mean limiting diffusion current i_d (in μA) as a function of cadmium(II) ion concentration c (in mM). A linear Cartesian coordinate system is selected based directly on the first-principles Ilkovič boundary value solution for transient convective-diffusive transport to an expanding spherical mercury droplet: i_d = 607 · n · D¹/² · m²/³ · t¹/⁶ · c. Under strictly constant capillary hydrostatic head and unvarying drop lifetime, the capillary characteristic m²/³ t¹/⁶ remains constant, requiring that diffusion current scale strictly linearly with analyte concentration and pass through the origin. Multi-point linear least-squares regression across all six experimental concentrations confirms rigorous linearity (R² = 1.000000) with an empirical Y-intercept b = -6.5672e-04 μA (indistinguishable from zero within experimental uncertainty). The empirical slope d(i_d)/dc = 6.3719 μA·mM⁻¹ enables direct, highly robust extraction of the diffusion coefficient D.

5. Method

  1. Adjust the mercury leveling bulb to establish a precise 50.0 cm column height. Collect 50 falling drops under deionized water into a tared weighing bottle to calibrate mercury mass flow rate m = 2.000 ± 0.005 mg/s.
  2. Prepare standardized Cd²⁺ solutions (0.200, 0.500, 1.000, 1.500, 2.000, 2.500 mM) in 0.1000 M KCl containing 0.002% w/v Triton X-100 to suppress polarographic streaming maxima.
  3. Transfer 25.0 mL of analyte into the jacketed H-cell at 25.00 °C. Bubble high-purity nitrogen for 15 minutes to purge dissolved oxygen, then divert gas flow over the solution surface.
  4. Configure the polarograph from E_initial = -0.200 V to E_final = -1.000 V vs SCE at a slow scan rate of 2.0 mV/s. Record the residual non-faradaic charging current prior to the reduction wave onset.
  5. Initiate the linear potential ramp. Record the sigmoidal polarographic wave. Measure the limiting plateau current (i_lim) between -0.80 V and -0.90 V and subtract the extrapolated residual current to obtain the true diffusion current (i_d).
  6. Verify half-wave potential invariance (E_1/2 = -0.600 V vs SCE). Plot mean diffusion current i_d against cadmium concentration c. Perform linear least-squares regression to extract slope d(i_d)/dc and calculate D using the Ilkovič equation.

6. Observations and calculation table

All gravimetric observations were recorded using a calibrated 4-decimal electronic analytical balance enclosed within a draft shield chamber. Volumetric aliquots were dispensed from a Class A burette. Residuals are computed relative to the ordinary least-squares regression line.

Table 1. Seven simulated gravimetric observations; x and y represent dispensed volume and net mass metrics.
No.Cadmium(II) Concentration c (mM)Mean Limiting Diffusion Current i_d (μA)Cadmium(II) Concentration c (mM)Mean Limiting Diffusion Current i_d (μA)Fitted y (g)Residual (g)
10.21.27000.20001.27001.2737-0.0037
20.53.19000.50003.19003.18530.0047
31.06.37001.00006.37006.3713-0.0013
41.59.56001.50009.56009.55730.0027
52.012.74002.000012.740012.7432-0.0032
62.515.93002.500015.930015.92920.0008

Download the sample observations (CSV) ↓

7. Worked calculation for one observation

1. Raw Experimental Observables & Operating Polarographic Coordinates: Cadmium(II) ion concentration: c = 0.200 mM = 2.0000e-04 mol·L⁻¹ in 0.100 M KCl supporting electrolyte. Measured mean limiting diffusion current: i_d = 1.27 μA. Half-wave potential: E_1/2 = -0.600 V vs SCE (invariance across all c coordinates verifies ideal reversibility). 2. Calibrated Capillary Characteristics of the Dropping Mercury Electrode (DME): Mercury column height: h = 50.0 cm. Mercury mass flow rate: m = 2.000 mg·s⁻¹ (measured via gravimetric collection of 50 drops under water). Individual drop lifetime: t = 3.500 s (measured via optical photodiode chronometer). Capillary constant: m²/³ t¹/⁶ = (2.000)²/³ × (3.500)¹/⁶ = 1.95598 mg²/³·s⁻¹/². 3. Multi-Point Ilkovič Linear Least-Squares Regression: Linearized theoretical equation: i_d = m_slope · c + b_int. Regression slope: m = 6.3719 μA·mM⁻¹. Regression Y-intercept: b = -6.6000e-04 μA (consistent with zero diffusion current at zero concentration within ±0.005 μA). Coefficient of determination: R² = 0.9999996. 4. Extraction of Aqueous Cadmium(II) Diffusion Coefficient D: Ilkovič constant prefactor = 607 · n · m²/³ · t¹/⁶ = 607 × 2 × 1.95598 = 2374.56 μA·mM⁻¹·(cm²·s⁻¹)⁻¹/². Diffusion coefficient: D = [ m_slope / prefactor ]² = [ 6.3719 / 2374.56 ]² = 7.2007e-06 cm²·s⁻¹. Scaled diffusion coefficient: D = 7.20 × 10⁻⁶ cm²·s⁻¹. Accepted electroanalytical reference benchmark at 25.00 °C: D = 7.20 × 10⁻⁶ cm²·s⁻¹ (percentage difference = 0.0102%).

8. Graph, gradient and final calculation

-0.9-0.13.80.68.61.413.42.118.12.8Cadmium(II) Concentration c (mM)Mean Limiting Diffusion Current i_d (μA)V = 0.2 mL, m = 1.27 gV = 0.5 mL, m = 3.19 gV = 1.0 mL, m = 6.37 gV = 1.5 mL, m = 9.56 gV = 2.0 mL, m = 12.74 gV = 2.5 mL, m = 15.93 g

Circular markers represent empirical gravimetric observations. The solid line represents the ordinary least-squares linear regression fit.

x̄ = 1.2833 mL; ȳ = 8.1767 g
Σ(x − x̄)² = 3.9083 mL²
Σ(x − x̄)(y − ȳ) = 24.9037 g·mL
m = Σ(x − x̄)(y − ȳ) / Σ(x − x̄)² = 6.37194030 g/mL
b = ȳ − mx̄ = -6.5672e-04 g
Ilkovič Linear Fit: i_d = 6.372·c + -6.5672e-04 μA; Cadmium Diffusion Coefficient D = 7.20 × 10⁻⁶ cm²·s⁻¹; Half-Wave Potential E_1/2 = -0.600 V vs SCE

9. Uncertainty and comparison

Percentage difference = |estimate − theoretical value| / |theoretical value| × 100 = 0.0102%.

For n = 7 gravimetric observations, Σresidual² = 5.6119e-05; s² = Σresidual²/(n − 2). Gradient standard error = √[s²/Σ(x − x̄)²] = 0.00189. Intercept standard error = √[s²(1/n + x̄²/Σ(x − x̄)²)] = 0.00287.

Propagating the fit parameters yields an experimental density standard error of 0.00428 10⁻⁶ cm²·s⁻¹, with an ideal coefficient of determination R² = 1.000000.

In the high-precision determination of heavy metal diffusion coefficients via classical polarography and the Ilkovič equation, metrological uncertainties and physical electroanalytical boundary conditions are rigorously quantified across capillary, gravimetric, potentiostatic, and chemical domains: (1) Mercury Mass Flow Rate Gravimetry: Mercury flow rate m is calibrated by collecting 50 successive drops under deionized water into a tared weighing bottle on a calibrated 5-decimal analytical balance. The certified mass flow rate is m = 2.000 ± 0.005 mg·s⁻¹, yielding a relative flow rate uncertainty u(m)/m = 0.25%. Hydrostatic head is fixed at 50.0 ± 0.1 cm using a cathetometer. (2) Drop-Time Optical Chronometry: Drop lifetimes are monitored using an optical infrared photodiode gate linked to a digital chronometer with microsecond timebase. For t_drop = 3.500 s, cycle-to-cycle repeatability across 100 drops yields a standard uncertainty u(t) = 0.005 s (0.14%). (3) Maxima Suppression & Capillary Cleanliness: Addition of exactly 0.002% w/v Triton X-100 completely suppresses Marangoni streaming maxima. The residual non-faradaic charging current is measured between -0.20 V and -0.40 V and subtracted from the total limiting current to isolate pure faradaic diffusion current i_d with a baseline subtraction uncertainty u(i_base) = 0.02 μA. (4) Multi-Point Least-Squares Regression & Uncertainty Propagation: The empirical Ilkovič slope m = 6.3719 μA·mM⁻¹ is evaluated by ordinary least-squares regression over six discrete concentration coordinates (0.200 to 2.500 mM). The standard error of the slope is s_m = 0.0035 μA·mM⁻¹ (relative uncertainty 0.055%), yielding a coefficient of determination R² = 0.9999996. (5) Combined Uncertainty Budget: Because D is proportional to [m / (m_Hg²/³ t¹/⁶ c)]², standard uncertainties propagate as u_r(D) = √[ 4(u(m)/m)² + (16/9)(u(m_Hg)/m_Hg)² + (1/9)(u(t)/t)² + 4(u(c)/c)² ] = 0.39%. For an empirical diffusion coefficient D = 7.20 × 10⁻⁶ cm²·s⁻¹, the expanded uncertainty at the 95% confidence level (coverage factor k = 2) is U(D) = ±0.06 × 10⁻⁶ cm²·s⁻¹ (0.8%), confirming sub-0.5% experimental agreement with accepted electroanalytical reference benchmarks.

10. Result and discussion

7.20 = 7.20 10⁻⁶ cm²·s⁻¹
CIPM standard reference value: 7.20 10⁻⁶ cm²·s⁻¹

The experimental polarograms provide profound insights into convective-diffusive mass transport and electrocapillary interfacial phenomena. The observed half-wave potential E_1/2 = -0.600 V vs SCE remained completely stationary across the entire 12.5-fold concentration range from 0.200 mM to 2.500 mM. According to the Heyrovský-Ilkovič equation, E_1/2 = E°' - (RT/2nF)·ln(D_Cd²⁺/D_Cd(Hg)), which depends solely on the intrinsic thermodynamics of cadmium amalgam formation and is independent of concentration and capillary geometry. The logarithmic analysis of the rising wave (plotting E vs log₁₀[i / (i_d − i)]) yielded a slope of -29.6 mV/decade, in exact agreement with the theoretical Nernstian prediction -59.2 mV / n for a two-electron reduction (n = 2), confirming that electron transfer is fast and thermodynamically reversible. The unique advantages of the dropping mercury electrode were vividly demonstrated by the remarkable reproducibility of the limiting diffusion plateau. Unlike solid electrodes, where reaction products deposit irreversibly on the surface causing time-dependent electrode fouling and history effects, each mercury droplet exposes a pristine, fresh liquid metal surface. When the drop detaches, the cadmium amalgam formed within the droplet is physically carried away into the bottom mercury pool, resetting the working interface to zero every 3.50 seconds. The incorporation of 0.002% w/v Triton X-100 was crucial in eliminating polarographic streaming maxima. Without surfactant, non-uniform current densities between the shielded neck and unshielded bottom pole of the drop create interfacial tension gradients (the Marangoni effect) that induce violent vortex streaming, surging the current artificially by 300% to 500%. The adsorbed Triton X-100 monolayer creates an elastic surface pressure that completely arrests tangential slip, restoring quiescent Fickian diffusion. The 0.100 M KCl supporting electrolyte suppressed the migration transference number of Cd²⁺ to < 0.5%, ensuring that mass transport occurred purely via molecular diffusion. The extracted diffusion coefficient D = (7.20 ± 0.06) × 10⁻⁶ cm²·s⁻¹ matches the Stokes-Einstein prediction for a hydrated divalent metal cation in water (hydrodynamic radius r_h ≈ 0.34 nm), validating the enduring authority of the Ilkovič equation in physical electrochemistry.

11. Precautions and limitations

['Classical DC polarography is limited to analyte concentrations above 10 μM due to non-faradaic capacitive charging currents (i_c ∝ m²/³ t⁻¹/³).', 'Capillary bore clogging by particulate debris or mercury oxidation can alter the capillary characteristic m²/³ t¹/⁶, requiring re-calibration.', 'Mercury cannot be utilized at potentials positive of +0.25 V vs SCE due to the anodic dissolution of metallic mercury (2 Hg + 2 Cl⁻ → Hg₂Cl₂ + 2 e⁻).'] The Ilkovič equation for the mean diffusion current at a dropping mercury electrode is i_d = 607 · n · D¹/² · m²/³ · t¹/⁶ · c, where n = 2 for Cd²⁺, m = 2.000 mg/s is the calibrated mercury mass flow rate, and t = 3.500 s is the drop lifetime. The capillary constant m²/³ t¹/⁶ equals 1.956 mg²/³·s⁻¹/². Across cadmium concentrations from 0.200 mM to 2.500 mM, the half-wave potential remains strictly invariant at E_1/2 = -0.600 ± 0.002 V vs SCE, confirming reversible two-electron amalgam formation. Linear least-squares regression of mean diffusion current i_d against cadmium concentration c yields a slope of 6.3719 μA·mM⁻¹ with R² = 0.9999996, extracting an aqueous Cd²⁺ diffusion coefficient D = 7.20 × 10⁻⁶ cm²·s⁻¹, in complete agreement with authoritative electrochemical literature benchmarks.

12. Conclusion

The aqueous diffusion coefficient of the cadmium(II) ion in 0.1000 M KCl at 25.00 °C was successfully determined as D = (7.20 ± 0.06) × 10⁻⁶ cm²·s⁻¹ using classical DC polarography at a dropping mercury electrode. The experimental limiting diffusion currents exhibited exceptional linear adherence to the Ilkovič equation (R² = 1.000000) across the 0.200 to 2.500 mM concentration range, with an invariant half-wave potential E_1/2 = -0.600 V vs SCE confirming ideal Nernstian reversibility. The extracted diffusion coefficient agrees with accepted literature benchmarks within 0.0102% relative deviation, establishing an authoritative experimental verification of convective-diffusive transport to expanding spherical liquid electrodes.