Diagnostic Cyclic Voltammetry and Mass Transport Kinetics via the Randles-Ševčík Equation

Determine the aqueous diffusion coefficient of the ferricyanide anion [Fe(CN)₆]³⁻ using cyclic voltammetry at a glassy carbon electrode across scan rates from 10.0 mV/s to 200.0 mV/s via the Randles-Ševčík equation. Explore interactive analytical gravimetry, Class A volumetric glassware and density determination.

MEASURE · CALIBRATE · ANALYSE

Your virtual chemistry laboratory.

PEAK CATHODIC CURRENT I_PC—
MEASUREMENT STATUS

Ready to dispense 50.0 mV·s⁻¹ of liquid.

50.0 mV·s⁻¹
Working Electrode: Polished Glassy Carbon Disk (d = 3.00 mm, Geometric Area A = 0.07069 cm²)

Adjust ν_scan V, then weigh the dispensed aliquot.

Mass observations are determined by the high-precision thermodynamic equation of state for pure liquid water. Readings reflect thermal equilibrium and analytical balance force restoration.

Physical chemistry assumptions and metrological limits

['The simple Randles-Ševčík equation assumes planar semi-infinite linear diffusion; at extremely slow scan rates (< 1 mV/s), spherical edge diffusion at microelectrodes causes positive deviations.', 'At very high scan rates (> 1000 mV/s), finite heterogeneous electron transfer kinetics (quasi-reversibility) and uncompensated cell resistance (iR_u) induce peak broadening and current attenuation.', 'Alumina polishing particles can become embedded in softer electrode matrices if excessive polishing pressure is applied, altering the electroactive surface area.'] The Randles-Ševčík equation for a reversible soluble redox couple at 25.00 °C (298.15 K) is i_pc = (2.686 × 10⁵) · n³/² · A · c · D¹/² · ν¹/², where n = 1, A = 0.07069 cm² (3.00 mm diameter glassy carbon disk), and c = 5.000 × 10⁻⁶ mol·cm⁻³ (5.000 mM K₃[Fe(CN)₆] in 1.000 M KNO₃). Across scan rates from 10.0 mV/s to 200.0 mV/s, peak cathodic currents increase from 26.18 μA to 117.06 μA while peak potential separation remains strictly constant at ΔE_p = 59.2 ± 0.5 mV. Linear least-squares regression of i_pc against ν¹/² yields a slope of 261.74 μA·(V·s⁻¹)⁻¹/² with R² = 0.99999999, extracting an aqueous ferricyanide diffusion coefficient D = 7.60 × 10⁻⁶ cm²·s⁻¹, in complete agreement with authoritative electrochemical literature benchmarks.

Analytical chemistry apparatus & tools.

Digital Potentiostat / Galvanostat Workstation

Precision bipotentiostat with sub-picoampere current resolution, 16-bit DAC potential ramps, and active IR-drop compensation.

Glassy Carbon Working Electrode (GC)

High-density vitreous carbon disk embedded in Teflon shroud (diameter 3.00 ± 0.01 mm, area A = 0.07069 cm²) mirror-polished with 0.05 μm alumina slurries.

Saturated Calomel Reference Electrode (SCE)

Double-junction reference electrode (+0.2412 V vs SHE at 25 °C) featuring a porous Vycor frit that establishes stable reference potential without chloride contamination.

Platinum Mesh Auxiliary (Counter) Electrode

Coiled platinum wire gauze (surface area > 2.0 cm²) facilitating unconstrained counter-current flow without polarizing the cell.

Thermostated Water-Jacketed Electrochemical Cell

Five-neck glass cell with water jacket maintained at 25.00 ± 0.02 °C, equipped with an argon bubbling tube and gas blanket head.

High-Purity Argon Purge & Blanket Gas System

Argon deoxygenation line with presaturator scrubbers to deoxygenate electrolyte solutions for 15 minutes, removing electroactive dissolved O₂.

FOLLOW THE METHOD

A reason behind every step.

Determine the aqueous diffusion coefficient of the ferricyanide anion [Fe(CN)₆]³⁻ using cyclic voltammetry at a glassy carbon electrode across scan rates from 10.0 mV/s to 200.0 mV/s via the Randles-Ševčík equation. Suggested independent volumetric settings: 10.0, 25.0, 50.0, 100.0, 150.0, 200.0 mV·s⁻¹.

01

Working Electrode Polishing & Microscopic Inspection

Polish the glassy carbon electrode on microcloth pads using aqueous alumina slurries (1.0 μm, 0.3 μm, and 0.05 μm successively) in figure-eight patterns. Sonicate in deionized water and ethanol for 5 minutes each to remove polishing particulates.

02

Electrolyte Formulation & Supporting Salt Addition

Prepare 50.0 mL of standardized 5.000 mM K₃[Fe(CN)₆] in 1.000 M KNO₃ supporting electrolyte using degassed, triple-distilled deionized water.

03

Cell Assembly & Argon Deoxygenation

Assemble the three-electrode cell in the 25.00 °C jacketed vessel. Bubble high-purity argon through the solution for 15 minutes to purge dissolved oxygen, then divert argon to create a continuous inert blanket over the headspace.

04

Potential Window Selection & Baseline Verification

Configure the potentiostat between initial potential E_i = +0.600 V and switching potential E_s = -0.100 V vs SCE. Execute a blank run in 1.0 M KNO₃ to verify that residual background currents remain below 0.1 μA.

05

Multi-Scan Rate Cyclic Voltammetric Acquisition

Record cyclic voltammograms across scan rates ν = 10.0, 25.0, 50.0, 100.0, 150.0, and 200.0 mV/s. Record the peak cathodic current (i_pc), peak anodic current (i_pa), and peak potential separation (ΔE_p = E_pa - E_pc).

06

Randles-Ševčík Diagnostic Regression & Diffusion Extraction

Verify reversibility criteria (ΔE_p ≈ 59 mV, i_pa/i_pc ≈ 1.0). Plot peak cathodic current i_pc against ν¹/² (in (V/s)¹/²). Perform linear least-squares regression to extract the slope and compute the diffusion coefficient D using the Randles-Ševčík equation.

YOUR NOTEBOOK · 0 OBSERVATIONS

Turn measurements into quantitative chemical insight.

See a complete example report ↗

Points are your recorded gravimetric observations. The regression line validates the linear mass-volume proportionality and extracts density.

ORDINARY LEAST-SQUARES FIT · m = ρ·V + m₀
i_{pc} = 0.4463 \cdot n F A c \left(\frac{n F \nu D}{R T}\right)^{1/2} \implies i_{pc} = 261.74 \cdot \nu^{1/2} \implies D = 7.60 \times 10^{-6}\text{ cm}^2\cdot\text{s}^{-1}
#Voltammetric Potential Scan Rate ν (mV·s⁻¹)Peak Cathodic Current i_pc (μA)Square Root of Scan Rate ν¹/² (V¹/²·s⁻¹/²)Peak Cathodic Current i_pc (μA)Remove

No readings yet. Dispense and weigh a volumetric aliquot to record it.

Calculate it yourself.

CHECK YOUR UNDERSTANDING

Explain what the gravimetric measurements mean.

What is the theoretical peak potential separation (ΔE_p = |E_pa - E_pc|) for a reversible 1-electron redox reaction at 25 °C?

What does a strictly linear plot of peak current i_pc versus the square root of scan rate (ν¹/²) indicate?

Why is an inert supporting electrolyte (e.g. 1.0 M KNO₃) added in large excess relative to the electroactive analyte?

In the Randles-Ševčík equation i_pc = 0.4463 · n F A c · √(n F ν D / R T), what happens to peak current if electrode diameter is doubled?