Biomedical Engineering / Medical Force Sensor
FREE TO EXPLORE
LABORATORY 02 · BIOMEDICAL ENGINEERING

Strain-Gauge Cantilever Beam Force Sensor & Sensitivity Calibration

Measure the differential voltage output of a full-bridge foil strain gauge on a medical force sensor as a function of applied tip load to evaluate gauge factor and flexural sensitivity. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.

BIOINSTRUMENTATION · BIOSIGNALS · BIOMECHANICS · IMAGING

Your virtual biomedical engineering laboratory.

BRIDGE DIFFERENTIAL OUTPUT V_OUT—
PHYSIOLOGICAL & TEST CELL TELEMETRY

Ready to sample with Force F = 3.0 N.

3.0 N
Beam material: 6061-T6 Aluminum alloy (E = 70 GPa)

Adjust Force F, then execute biomedical trial.

Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.

Biomedical assumptions and boundary condition limits

The linear model applies strictly within the elastic limit of the aluminum beam (strain ε < 1000 με). Evaluated on a medical-grade aluminum alloy cantilever beam (E = 70 GPa) with four 350 Ω foil strain gauges (GF = 2.10) wired in a full-bridge circuit energized by a 10.0 V DC reference. Sensitivity S_f is evaluated across loads from 1.0 N to 6.0 N.

Biomedical apparatus & physiological sensors.

Instrumented Cantilever Load Cell Beam

Provides a precision-machined aluminum alloy flexure beam with four bonded 350 Ω foil strain gauges.

Full Wheatstone Bridge Conditioning Board

Arranges two upper tensile and two lower compressive gauges in push-pull bridge configuration.

Class F1 Precision Calibration Deadweight Set

Applies traceable vertical calibration forces from 1.0 N to 6.0 N (±0.01% tolerance).

Stable Low-Noise Bridge Power Supply

Provides highly regulated 10.000 V DC excitation with < 50 μV ripple.

Precision Instrumentation Amplifier & DMM

Amplifies microvolt bridge output and digitizes differential signals with 6.5-digit precision.

Rigid Clamped Mounting Fixture

Prevents base rotation or compliance slippage during tip load application.

FOLLOW THE METHOD

A reason behind every step.

Measure the differential voltage output of a full-bridge foil strain gauge on a medical force sensor as a function of applied tip load to evaluate gauge factor and flexural sensitivity. Standardized parameter settings: 1.0, 2.0, 3.0, 4.0, 5.0, 6.0 N.

01

Bridge Zero Balancing and Shunt Calibration

Secure the cantilever base into the rigid clamping block and adjust the bridge balance circuit until unloaded differential output reads 0.0000 mV.

02

Excitation Voltage Verification

Measure bridge input excitation voltage at terminals with calibrated DMM; confirm steady 10.000 V reference.

03

Initial Load Increment Application

Gently place the 1.000 N calibration weight onto the knife-edge tip loading hanger without dynamic shock.

04

Settling and Telemetry Acquisition

Allow 15 seconds for viscoelastic relaxation and mechanical vibrations to settle; record stabilized differential bridge voltage V_out in millivolts.

05

Stepwise Loading Progression

Progressively add calibration masses to achieve loads of 2.0, 3.0, 4.0, 5.0, and 6.0 N, logging stabilized output voltage at each load.

06

Sensitivity Regression and Linearity Verification

Plot differential voltage V_out against applied force F, performing least-squares linear regression to determine force sensitivity S_f.

YOUR NOTEBOOK · 0 OBSERVATIONS

Turn measurements into biomedical engineering insight.

See a complete example report ↗

Points are your recorded observations. The solid regression line validates Medical Force Sensor mathematical scaling.

LINEAR REGRESSION FIT · y = mx + b
V_out = S_f · F = 2.4500 · F
#Applied Normal Force F (N)Bridge Differential Output V_out (mV)Applied force F (N)Differential output V_out (mV)Remove

No readings yet. Execute a biomedical trial and record it.

Calculate it yourself.

CHECK YOUR UNDERSTANDING

Explain what the biomedical metrics mean.

What is the typical gauge factor (GF) for standard metallic foil strain gauges?

Why are two gauges mounted on top and two on the bottom of a cantilever beam load cell?

If applied force is 4.0 N and bridge output is 9.8000 mV, what is the force sensitivity S_f?

Where is the surface bending strain maximum in an end-loaded cantilever beam?