Anatomical Transtibial Residual Limb Mannequin
Provides rigid skeletal core (tibia/fibula) surrounded by elastomeric soft-tissue simulating muscle and skin.
Map the interface contact pressure at the patellar-tendon bar of a patellar-tendon-bearing (PTB) prosthetic socket as a function of axial limb weight-bearing load. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.
Ready to sample with Axial Load W = 400.0 N.
Adjust Axial Load W, then execute biomedical trial.
Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.
The linear sensitivity model applies to static axial loading; dynamic ambulation introduces coupled shear stresses and moment loads. Evaluated on an anatomically contoured robotic residual limb mannequin with silicone soft tissue surrogate (Shore 20A durometer) fitted inside a carbon-fiber PTB prosthetic socket. Instrumented with calibrated thin-film piezoresistive pressure sensors at the patellar-tendon bar. Linear sensitivity k_soc = 0.2200 kPa/N across loads from 200 N to 700 N.
Provides rigid skeletal core (tibia/fibula) surrounded by elastomeric soft-tissue simulating muscle and skin.
Custom-laminated below-knee socket with inward patellar-tendon weight-bearing indentation bar.
Matrix of flexible sensors (thickness 0.15 mm) measuring local interface pressure up to 250 kPa.
Applies controlled vertical limb loads W from 200 N to 700 N with closed-loop force control.
Scans 96 pressure matrix sensing cells at 100 Hz with 8-bit gray-scale spatial mapping.
Applies uniform hydrostatic pressure steps to calibrate piezoresistive array prior to limb insertion.
Map the interface contact pressure at the patellar-tendon bar of a patellar-tendon-bearing (PTB) prosthetic socket as a function of axial limb weight-bearing load. Standardized parameter settings: 200.0, 300.0, 400.0, 500.0, 600.0, 700.0 N.
Place thin-film sensor array inside pneumatic bladder calibration chamber and apply 50, 100, and 150 kPa pressures to generate cell-by-cell calibration lookup tables.
Affix calibrated sensor matrix securely along the patellar-tendon bar and medial tibial flare of the residual limb mannequin using skin-safe adhesive film.
Carefully don a 6 mm silicone gel prosthetic liner and seat the residual limb into the carbon-fiber PTB socket, verifying anatomical alignment.
Position the prosthetic assembly under the vertical loading ram and apply a 50 N tare load to bed down the elastomeric liner.
Increment axial vertical load through 200, 300, 400, 500, 600, and 700 N, recording stabilized mean patellar-tendon contact pressure P_ptb at each step.
Plot patellar-tendon interface pressure P_ptb against total axial load W and compute the socket load transfer constant k_soc via linear regression.
Points are your recorded observations. The solid regression line validates Prosthetic Socket Interface Pressure mathematical scaling.
| # | Applied Axial Limb Load W (N) | Patellar-Tendon Interface Pressure P_ptb (kPa) | Axial limb load W (N) | Patellar tendon interface pressure P_ptb (kPa) | Remove |
|---|
No readings yet. Execute a biomedical trial and record it.