ISO 5840 Steady-Flow Heart Valve Test Rig
Provides straight upstream and anatomical sinus of Valsalva downstream flow conduits with pressure taps.
Measure the transvalvular pressure gradient ΔP across a 25 mm bileaflet mechanical aortic heart valve under steady forward flow and evaluate the quadratic Bernoulli loss coefficient k_v. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.
Ready to sample with Velocity v = 1.4 m/s.
Adjust Velocity v, then execute biomedical trial.
Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.
Steady forward flow does not simulate transient valve opening/closing leaflet dynamics or dynamic regurgitation phases. Evaluated on a 25 mm bileaflet pyrolytic carbon mechanical aortic valve mounted in an acrylic aortic root chamber meeting ISO 5840 standards. Perfusion fluid is 36% glycerol blood analog (density ρ = 1055 kg/m³, viscosity 3.6 cP at 37 °C). Velocities v = 1.0, 1.2, 1.4, 1.6, 1.8, 2.0 m/s correspond to v² = 1.00, 1.44, 1.96, 2.56, 3.24, 4.00 m²/s². Quadratic Bernoulli slope k_v = 3.8500 mmHg/(m²/s²).
Provides straight upstream and anatomical sinus of Valsalva downstream flow conduits with pressure taps.
Clinical prosthesis with dual semicircular pyrolytic carbon leaflets and Dacron sewing cuff.
Supplies stable, continuous non-pulsatile forward flow from 5 to 35 L/min.
Measures transvalvular pressure drop ΔP with pressure taps located 1 diameter upstream and 3 diameters downstream.
Measures volumetric forward flow rate Q with 1% accuracy and sub-millisecond response.
Maintains fluid density ρ = 1055 kg/m³ and blood-analog viscosity 3.6 cP.
Measure the transvalvular pressure gradient ΔP across a 25 mm bileaflet mechanical aortic heart valve under steady forward flow and evaluate the quadratic Bernoulli loss coefficient k_v. Standardized parameter settings: 1.0, 1.2, 1.4, 1.6, 1.8, 2.0 m/s.
Mount 25 mm mechanical valve securely in acrylic test chamber, ensuring leaflets open symmetrically without sewing ring impingement.
Fill system with 36% glycerol-water blood analog, purge all trapped air from pressure lines, and heat reservoir to 37.0 °C.
Open transducer bypass valve at zero flow to establish precise 0.00 mmHg baseline differential pressure.
Adjust centrifugal pump to establish transvalvular jet velocity v = 1.0 m/s (verified by flowmeter Q and effective orifice area) and record steady pressure drop ΔP.
Increment jet velocity through 1.2, 1.4, 1.6, 1.8, and 2.0 m/s, recording stabilized pressure gradient ΔP over 30 seconds at each step.
Plot transvalvular pressure drop ΔP against v², compute the slope k_v via linear regression through the origin, and calculate effective orifice area EOA.
Points are your recorded observations. The solid regression line validates Heart Valve Steady Pressure Drop mathematical scaling.
| # | Transvalvular Jet Velocity v (m/s) | Transvalvular Pressure Drop ΔP (mmHg) | Square of transvalvular velocity v² (m²/s²) | Transvalvular pressure drop ΔP (mmHg) | Remove |
|---|
No readings yet. Execute a biomedical trial and record it.