Clinical Hemodialysis Delivery System Test Bench
Provides calibrated peristaltic blood pump and heated, degassed dialysate preparation system.
Evaluate small-solute (urea) clearance K_u of a clinical hollow-fiber hemodialyzer across varying blood flow rates Q_b at fixed dialysate flow rate Q_d. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.
Ready to sample with Blood Flow Q_b = 250.0 mL/min.
Adjust Blood Flow Q_b, then execute biomedical trial.
Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.
The linear clearance-flow model applies across standard clinical blood flows (150 to 400 mL/min); at extreme flows (> 500 mL/min), clearance plateaus asymptotically. Evaluated on a 1.8 m² high-flux polysulfone hollow-fiber hemodialyzer (Fresenius FX80, mass transfer-area coefficient K_0·A = 950 mL/min for urea). Constant counter-current dialysate flow Q_d = 500 mL/min at 37.0 °C with zero ultrafiltration (Q_f = 0). Linear blood flow clearance slope m_c = 0.5800 with baseline K_u0 = 42.0000 mL/min across blood flow rates Q_b from 150 to 400 mL/min.
Provides calibrated peristaltic blood pump and heated, degassed dialysate preparation system.
Contains ~12,000 semipermeable hollow fibers with 1.8 m² effective mass transfer surface area.
Provides standardized blood substitute containing physiological electrolytes and clinical urea concentration.
Prepares bicarbonate-buffered dialysate at 37.0 °C and meters counter-current flow through shell side.
Measures blood inlet (C_bi) and outlet (C_bo) urea concentrations using the urease-GLDH method.
Continuously measures blood compartment flow Q_b and dialysate flow Q_d with 1 mL/min accuracy.
Evaluate small-solute (urea) clearance K_u of a clinical hollow-fiber hemodialyzer across varying blood flow rates Q_b at fixed dialysate flow rate Q_d. Standardized parameter settings: 150.0, 200.0, 250.0, 300.0, 350.0, 400.0 mL/min.
Mount dialyzer vertically, prime blood side with 1 L heparinized saline at 150 mL/min to displace air, and invert to purge dialysate compartment.
Set dialysate delivery to 500 mL/min at 37.0 °C with bicarbonate conductivity 14.0 mS/cm and verify zero ultrafiltration transmembrane pressure.
Prepare synthetic blood reservoir with urea concentration C_bi = 25.0 mmol/L (150 mg/dL BUN) at 37.0 °C under gentle magnetic stirring.
Set blood pump to delivery rate Q_b = 150 mL/min, circulate counter-current to dialysate, and allow 5 minutes for concentration steady state.
Simultaneously sample inlet port (C_bi) and dialyzer outlet port (C_bo), assaying urea concentrations via spectrophotometric enzymatic assay to calculate clearance K_u.
Repeat sampling at Q_b = 200, 250, 300, 350, and 400 mL/min, plot urea clearance K_u versus Q_b, and determine flow sensitivity slope m_c.
Verify dialyzer performance against the theoretical logarithmic mean mass transfer relation K_0·A = [Q_b·Q_d / (Q_d - Q_b)] · ln[(1 - K/Q_d)/(1 - K/Q_b)].
Points are your recorded observations. The solid regression line validates Hemodialyzer Urea Clearance mathematical scaling.
| # | Blood Compartment Flow Rate Q_b (mL/min) | Dialytic Urea Clearance K_u (mL/min) | Blood flow rate Q_b (mL/min) | Urea clearance K_u (mL/min) | Remove |
|---|
No readings yet. Execute a biomedical trial and record it.