Precision Dual-Syringe Pulseless Perfusion Pump
Delivers pulseless volumetric flow rates from 0.1 to 20 mL/min with stepper-motor accuracy.
Determine the hydraulic flow resistance R_hyd of a precision microvascular capillary network perfusion model and verify Hagen-Poiseuille laminar resistance scaling. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.
Ready to sample with Flow Rate Q = 6.0 mL/min.
Adjust Flow Rate Q, then execute biomedical trial.
Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.
Hagen-Poiseuille law assumes a rigid, circular, straight tube; physiological blood vessels are compliant, branching, and tapered. Evaluated on a biomimetic PDMS microvascular capillary perfusion cartridge (lumen diameter 120 μm, length 40 mm, 8 parallel microchannels) perfused with physiological saline (viscosity μ = 0.0010 Pa·s) at 37.0 °C. Perfusion flow rates from 2.0 to 12.0 mL/min yield Reynolds numbers Re < 50 (strictly laminar). Hydraulic resistance R_hyd = 3.6000 kPa/(mL/min).
Delivers pulseless volumetric flow rates from 0.1 to 20 mL/min with stepper-motor accuracy.
Provides microfabricated parallel cylindrical conduits with optically transparent viewing windows.
Measures trans-vascular pressure drop ΔP from 0 to 50 kPa with 0.05% accuracy.
Maintains perfusate fluid temperature stable within ±0.1 °C to fix fluid viscosity.
Eliminates microscopic air bubbles that would lodge in microchannels and alter hydraulic resistance.
Validates actual volumetric delivery rate by timed gravimetric fluid collection.
Determine the hydraulic flow resistance R_hyd of a precision microvascular capillary network perfusion model and verify Hagen-Poiseuille laminar resistance scaling. Standardized parameter settings: 2.0, 4.0, 6.0, 8.0, 10.0, 12.0 mL/min.
Prime the microvascular flow loop with degassed physiological saline at 37.0 °C, ensuring zero air bubbles remain in the tubing or differential pressure ports.
With the perfusion pump halted, open the bypass valve to equalize pressure ports and zero the differential pressure transducer to exactly 0.000 kPa.
Set the dual-syringe infusion pump to deliver 2.0 mL/min and close the bypass valve, allowing 2 minutes for pressure to achieve steady state.
Record 100 consecutive differential pressure readings over 30 seconds and calculate mean pressure drop ΔP in kilopascals.
Increment pump delivery rate through 4.0, 6.0, 8.0, 10.0, and 12.0 mL/min, logging stabilized pressure drop at each flow rate.
Plot trans-vascular pressure drop ΔP versus flow rate Q and determine hydraulic resistance R_hyd from the slope of the linear fit through the origin.
Points are your recorded observations. The solid regression line validates Microvascular Hydraulic Resistance mathematical scaling.
| # | Perfusion Volumetric Flow Rate Q (mL/min) | Perfusion Pressure Drop ΔP (kPa) | Perfusion flow rate Q (mL/min) | Perfusion pressure drop ΔP (kPa) | Remove |
|---|
No readings yet. Execute a biomedical trial and record it.