Magnetically Levitated Centrifugal Blood Pump Cartridge
Features bearingless MagLev impeller (diameter 36 mm) driven by brushless DC stator coils.
Determine the shut-off pressure head scaling constant k_N of a magnetically levitated continuous-flow rotary blood pump (LVAD) across operating speeds. Explore interactive physiological instrumentation simulation, biomechanics, and biomedical engineering telemetry.
Ready to sample with Speed N = 3.0 k-RPM.
Adjust Speed N, then execute biomedical trial.
Biomedical engineering telemetry and physiological states are evaluated rigorously using 64-bit IEEE 754 precision. Residuals reflect ordinary least-squares regression across successive observations. Continuous data structures support high-fidelity biological and physiological modeling.
Zero-flow shut-off testing evaluates the maximum static head; operational LVADs deliver 4 to 6 L/min where dynamic frictional losses reduce delivered head. Evaluated on a magnetically levitated continuous-flow centrifugal blood pump (impeller diameter 36 mm) circulating a blood-analog fluid (38% glycerol-water, density ρ = 1055 kg/m³, viscosity 3.6 cP at 37 °C). Speeds N = 2.0, 2.5, 3.0, 3.5, 4.0, 4.5 k-RPM correspond to N² = 4.00, 6.25, 9.00, 12.25, 16.00, 20.25 k-RPM². Shut-off head scaling constant k_N = 7.5000 mmHg/k-RPM².
Features bearingless MagLev impeller (diameter 36 mm) driven by brushless DC stator coils.
Provides compliant hydraulic flow circuit with precision pinch valve for zero-flow shut-off testing.
Measures pump inlet pressure P_in and outlet pressure P_out to determine pressure head H = P_out - P_in.
Verifies true zero flow (Q = 0.00 L/min) during shut-off head measurements.
Maintains closed-loop impeller rotational speed N within ±1 RPM of setpoint.
Maintains blood-analog fluid viscosity (3.6 cP) and density (1055 kg/m³).
Determine the shut-off pressure head scaling constant k_N of a magnetically levitated continuous-flow rotary blood pump (LVAD) across operating speeds. Standardized parameter settings: 2.0, 2.5, 3.0, 3.5, 4.0, 4.5 k-RPM.
Mount the centrifugal blood pump in the mock loop, prime with 38% glycerol blood-analog at 37.0 °C, and purge all microscopic air bubbles.
Energize active magnetic levitation coils and verify central impeller suspension without mechanical housing contact.
With the pump stopped (N = 0 RPM), open cross-bypass to confirm equal inlet and outlet pressure, zeroing differential transducer to 0.00 mmHg.
Close the downstream outflow valve to establish zero volumetric flow (Q = 0.00 L/min), set impeller speed to N = 2.0 k-RPM (2000 RPM), and log shut-off head H_0.
Increment rotor speed through 2.5, 3.0, 3.5, 4.0, and 4.5 k-RPM, recording stabilized shut-off pressure head H_0 at each speed step.
Plot shut-off head H_0 against N², determine pump scaling constant k_N via linear regression through the origin, and calculate maximum blade tip velocity.
Points are your recorded observations. The solid regression line validates Rotary Blood Pump Head-Flow mathematical scaling.
| # | Impeller Rotational Speed N (k-RPM) | Zero-Flow Shut-Off Head H₀ (mmHg) | Square of rotational speed N² (k-RPM²) | Shut-off pressure head H₀ (mmHg) | Remove |
|---|
No readings yet. Execute a biomedical trial and record it.